******        H O T F I X   5   F O R   A B B A / E   2 . 3 0 G         *******     
******                                                                  *******
*******************************************************************************
          

Installationprerequisite :  ABBA/E Software 2.30G 

Hotfix 5 contains all older Hotfixes for 2.30G 

Installation             :  - Reset the Rho Controller,
                              no control on, only mainswitch on
                            - Upload original Files from Version 2.30G
                              to your Harddisk (Backup for safety)
                            - Download the files to the Rho Controller
                            - Reset the Rho Controller 



*******************************************************************************
*************************       HOTFIX 5:       *******************************
*******************************************************************************


Related ticket           :  ----

Contents of the floppy   :  ABBAE\Ver230g\Fix5\Parallel\FLWMULTI.IRD
                            ABBAE\Ver230g\Fix5\Parallel\FRACK.IRD                            
                            ABBAE\Ver230g\Fix5\Parallel\HTURM.IRD 
                            ABBAE\Ver230g\Fix5\Parallel\INIT.IRD
                            ABBAE\Ver230g\Fix5\Parallel\PERMAN.IRD
                            ABBAE\Ver230g\Fix5\Parallel\FNEWGRIP.IRD
                            ABBAE\Ver230g\Fix5\PIC1\IQ_ABBAE.P2X
                            ABBAE\Ver230g\Fix5\DTA\*.DTA

NOTE: set all *.ird files in the Version.dat to 2307. 
      ( PERMAN and INIT are together as INIT Entry).





Description   HTURM/Perman  : - the position of the handlebuttons was not correct.
                              - the look command on a drive will not longer move
                                to the endposition from the drive. It stopped at the end
                                of the main-movement. 

 
Description   FLWDLTLP.DTA  : - wrong y-value under the INIT ENTRY.



*******************************************************************************


Beschreibung  HTURM/Perman  : - das anfahren der handles war nicht korrekt.
                              - Bei Look Kommand auf ein Laufwerk wird nicht laenger
                                bis zur Endposition des Laufwerkes gefahren. Es wird
                                nach der Hauptanfahrbewegung gestoppt.

 
Beschreibung  FLWDLTLP.DTA  : - falscher Y-Wert unterhalb des INIT Eintrag.

EMASS-Grau Storage Systems, 20.03.2000

*******************************************************************************
*************************       HOTFIX 4:       *******************************
*******************************************************************************


Related ticket           :  ----

Contents of the floppy   :  ABBAE\Ver230g\Fix4\Parallel\FLWMULTI.IRD
                            ABBAE\Ver230g\Fix4\Parallel\FRACK.IRD                            
                            ABBAE\Ver230g\Fix4\Parallel\HTURM.IRD 
                            ABBAE\Ver230g\Fix4\Parallel\INIT.IRD
                            ABBAE\Ver230g\Fix4\Parallel\PERMAN.IRD
                            ABBAE\Ver230g\Fix4\Parallel\FNEWGRIP.IRD
                            ABBAE\Ver230g\Fix4\PIC1\IQ_ABBAE.P2X
                            ABBAE\Ver230g\Fix4\DTA\*.DTA

NOTE: set all *.ird files in the Version.dat to 2307. 
      ( PERMAN and INIT are together as INIT Entry).



New included drive "FUJITSU DIANA4 ( M8100 )". File FLWDIAN4.DTA


Description   IQ_ABBAE   : - New PIC File for the RHO 3.0 with only one 
                             Input Card. 

Description   FNEWGRIP   : - change of the newgrip. The last movement ist now rectangular
                             VALUE in VERSION.DAT = 2307
                        	       

Description   HTURM      : - Fix for Hexatower 4 ( not in right position )


Description   FLWMULTI   : - on the second Newgrip apears a crash at
                             the wall, if the unit is not in the
                             same segment as the newgripstation.
                           - if the Bow in backposition signal was
                             missing at the second crash on a unit
                             the gripper was opening, and lost the
                             cartridge.
                           - If a crash apears on the Tower or Rack,
                             there was no move backward. The System have to
                             be reseted.
                           - Correction in Handling, GET Offset against PUT
                             KONFIG.DAT (Y_GET_LW) Example: 500,200
                             500 is for GET and PUT, when no second value
                             exists.
                             200 is a GET OFFSET
                           - G_BGL in *.DAT Files has an new extended function.
                             Look at the DTA.TXT in DTA Direktory
                           - Some *.DTA parameterdescriptions are changed 
                             Look at the DTA.TXT in DTA Direktory 
                           - DCI for Low Profile DLT is suppored ( Unittype  DG )

     INIT/PERMAN/FRACK   : - The DLT Tape will was moved (z) not so high as
                             before
                           - DCI for Low Profile DLT is suppored ( Unittype  DG )


*******************************************************************************

Neues Laufwerk "FUJITSU DIANA4 ( M8100 )". File FLWDIAN4.DTA


Beschreibung  IQ_ABBAE   : - Neues Pic für RHO 3.0 mit einer Eingangskarte 

Beschreibung  FNEWGRIP   : - rechtwinkliges anfahren beim greifen der Kassette.
                             Eintrag un VERSION.DAT = 2307 

Beschreibung  HTURM      : - Fix fuer Hexaturm 4 

Beschreibung   FLWMULTI  : - nach dem zweiten Ausrichen wurde an die
                             Verkleidung gefahren, wenn die Unit nicht
                             im gleichen Bereich wie Ausrichtsatation ist.
                           - bei fehlendem Bgel hinten Signal wurde beim
                             zweiten crash an der Unit die Kassette
                             losgelassen.
                           - Bei crash am Turm oder Rack wurde nicht
                             zurueckgefahren. Steuerung musste
                             resettet werden.
                           - Korrektur Handling, GET Versatz Laufwerk zu PUT
                             KONFIG.DAT (Y_GET_LW)  Bsp.: 500,200
                             500 wirkt auf GET und PUT
                             200 wirkt als GET Versatz
                           - DCI FUER LOW PROFILE WIRD UNTERSTUETZT ( LAUFERKSTYP DG )
     INIT/PERMAN/FRACK   : - Das DLT Tape wird beim GET nicht mehr so
                             hoch (Z) angehoben
                           - G_BGL in *.DAT Files hat eine erweiterte Funktion.
                             Mehr Details in der DTA.TXT im DTA Verzeichniss
                           - Einige *.DTA Parameterbeschreibungen wurden geaendert
                             Mehr Details in der DTA.TXT im DTA Verzeichniss
                           - DCI FUER LOW PROFILE WIRD UNTERSTUETZT ( LAUFERKSTYP DG )


EMASS-Grau Storage Systems, 11.11.99

*******************************************************************************
*************************       HOTFIX 3:       *******************************
*******************************************************************************
  

Related ticket           :  ----

Contents of the floppy   :  ABBAE\Ver230g\Fix3\Parallel\FLWMULTI.IRD
                            ABBAE\Ver230g\Fix3\Parallel\FRACK.IRD                            
                            ABBAE\Ver230g\Fix3\Parallel\INIT.IRD
                            ABBAE\Ver230g\Fix3\Parallel\PERMAN.IRD
                            ABBAE\Ver230g\Fix3\DTA\*.DTA

NOTE: set all *.ird files in the Version.dat to 2307. 
      ( PERMAN and INIT are together as INIT Entry).


Description   FLWMULTI   : - on the second Newgrip apears a crash at
                             the wall, if the unit is not in the
                             same segment as the newgripstation.
                           - if the Bow in backposition signal was
                             missing at the second crash on a unit
                             the gripper was opening, and lost the
                             cartridge.
                           - If a crash apears on the Tower or Rack,
                             there was no move backward. The System have to
                             be reseted.
                           - Correction in Handling, GET Offset against PUT
                             KONFIG.DAT (Y_GET_LW) Example: 500,200
                             500 is for GET and PUT, when no second value
                             exists.
                             200 is a GET OFFSET
                           - G_BGL in *.DAT Files has an new extended function.
                             Look at the DTA.TXT in DTA Direktory
                           - Some *.DTA parameterdescriptions are changed 
                             Look at the DTA.TXT in DTA Direktory 

     INIT/PERMAN/FRACK   : - The DLT Tape will was moved (z) not so high as
                             before

*******************************************************************************

Beschreibung   FLWMULTI  : - nach dem zweiten Ausrichen wurde an die
                             Verkleidung gefahren, wenn die Unit nicht
                             im gleichen Bereich wie Ausrichtsatation ist.
                           - bei fehlendem Bgel hinten Signal wurde beim
                             zweiten crash an der Unit die Kassette
                             losgelassen.
                           - Bei crash am Turm oder Rack wurde nicht
                             zurueckgefahren. Steuerung musste
                             resettet werden.
                           - Korrektur Handling, GET Versatz Laufwerk zu PUT
                             KONFIG.DAT (Y_GET_LW)  Bsp.: 500,200
                             500 wirkt auf GET und PUT
                             200 wirkt als GET Versatz
     INIT/PERMAN/FRACK   : - Das DLT Tape wird beim GET nicht mehr so
                             hoch (Z) angehoben
                           - G_BGL in *.DAT Files hat eine erweiterte Funktion.
                             Mehr Details in der DTA.TXT im DTA Verzeichniss
                           - Einige *.DTA Parameterbeschreibungen wurden geaendert
                             Mehr Details in der DTA.TXT im DTA Verzeichniss


EMASS-Grau Storage Systems, 25.02.99

*******************************************************************************
*************************       HOTFIX 2:       *******************************
*******************************************************************************


Related ticket           :  ----

Contents of the floppy   :  FLWMULTI.IRD
                            FRACK.IRD                            
                            INIT.IRD
                            PERMAN.IRD
                            VERSION.DAT

NOTE: set all *.ird files in the Version.dat to 2307. 
      ( PERMAN and INIT are together as INIT Entry).


Description              : - on the second Newgrip apears a crash at
                             the wall, if the unit is not in the
                             same segment as the newgripstation.
                           - if the Bow in backposition signal was
                             missing at the second crash on a unit
                             the gripper was opening, and lost the
                             cartridge.
                           - If a crash apears on the Tower or Rack,
                             there was no move backward. The System have to
                             be reseted.
                           - Correction in Handling, GET Offset against PUT
                             KONFIG.DAT (Y_GET_LW) Example: 500,200
                             500 is for GET and PUT, when no second value
                             exists.
                             200 is a GET OFFSET
                           - The DLT Tape will was moved (z) not so high as
                             before

*******************************************************************************

Beschreibung             : - nach dem zweiten Ausrichen wurde an die
                             Verkleidung gefahren, wenn die Unit nicht
                             im gleichen Bereich wie Ausrichtsatation ist.
                           - bei fehlendem Bgel hinten Signal wurde beim
                             zweiten crash an der Unit die Kassette
                             losgelassen.
                           - Bei crash am Turm oder Rack wurde nicht
                             zurueckgefahren. Steuerung musste
                             resettet werden.
                           - Korrektur Handling, GET Versatz Laufwerk zu PUT
                             KONFIG.DAT (Y_GET_LW)  Bsp.: 500,200
                             500 wirkt auf GET und PUT
                             200 wirkt als GET Versatz
                           - Das DLT Tape wird beim GET nicht mehr so
                             hoch (Z) angehoben


EMASS-Grau Storage Systems, 08.01.99


*******************************************************************************
*************************       HOTFIX 1:       *******************************
*******************************************************************************


Related ticket           :  ----

Contents of the floppy   :  FLWMULTI.IRD
                            FRACK.IRD                            
                            VERSION.DAT

NOTE: set all *.ird files in the Version.dat to 2307. 
      ( PERMAN and INIT are together as INIT Entry).


Description:  flwmulti   : - on the second Newgrip apears a crash at
                             the wall, if the unit is not in the
                             same segment as the newgripstation.
                           - if the Bow in backposition signal was
                             missing at the second crash on a unit
                             the gripper was opening, and lost the
                             cartridge.
              frack      : - If a crash apears on the Tower or Rack,
                             there was no move backward. The System have to
                             be reseted.

*******************************************************************************

Beschreibung: flwmulti   : - nach dem zweiten Ausrichen wurde an die
                             Verkleidung gefahren, wenn die Unit nicht
                             im gleichen Bereich wie Ausrichtsatation ist.
                           - bei fehlendem Bgel hinten Signal wurde beim
                             zweiten crash an der Unit die Kassette
                             losgelassen.
              frack      : - Bei crash am Turm oder Rack wurde nicht
                             zurueckgefahren. Steuerung musste
                             resettet werden.


EMASS-Grau Storage Systems, 08.01.99